Materials & Processes for Medical Devices (August 10- 13): Fatigue Strength Enhancement of Ti-6Al-4V ELI Femoral Distal Stems Using Low Plasticity Burnishing

6.3 Fatigue Strength Enhancement of Ti-6Al-4V ELI Femoral Distal Stems Using Low Plasticity Burnishing

Tuesday, August 11, 2009: 11:20 AM
Salon F (Hilton Minneapolis )
Mr. Douglas Hornbach , Lambda Technologies, Cincinnati, OH
Mr. Paul Prevey , Lambda Technologies, Cincinnati, OH
Mr. Don McNulty , DePuy Orthopaedics, Inc., Warsaw, IN
Mr. Jonathan Carr , DePuy Orthopaedics, Inc., Warsaw, IN
Mr. Paul Sade , DePuy Orthopaedics, Inc., Warsaw, IN
Amy Craft , DePuy Orthopaedics, Inc., Warsaw, IN
Total hip replacement surgery is often required to relieve pain brought on by damage from disease or fracture as well as provide an improved range of motion. Advanced technologies in hip implants have provided stronger, longer lasting devices that meet the ever-increasing demands of more active patients. Modular hip implants are becoming more popular because they allow the surgeon to interoperatively tailor the size of the implant to the patient's individual anatomy.

A customized low plasticity burnishing (LPB) process was developed and applied to

Ti-6Al-4V ELI distal stem tapers, used in a modular hip construct, to provide increased fatigue strength and improved tolerance to fretting damage. The LPB process produced a deep layer of compression while providing a superior surface finish. High cycle fatigue results obtained on LPB treated stems are compared to untreated stems. Residual stresses produced by the LPB process were characterized via x-ray diffraction methods on the treated taper region of the stems. Residual stress results are shown for both LPB treated and untreated parts.