Corrosion Behavior of Cardiovascular Stents

N. D. Munroe, K. H. Wu, A. Datye, W. Haider, Florida International University, Miami, FL

Most alloys are electropolished prior to implantation and are immediately protected by the formation of passivating metal oxides. Nevertheless, tissue fluid present in the human body presents a very corrosive environment to metals once implanted. This is exacerbated by the effects of strain, fatigue, abrasion or wear, which may rupture the passive film exposing the underlying metal. Cyclic Potentiodynamic polarization studies were conducted to compare the corrosion resistance of nitinol and stainless steel cardiovascular stents. In an atempt to simulate their corrosion behavior under body conditions, a Phosphate Buffered Saline (PBS) solution at 37 oC was employed utilizing the ASTM F2129-01, Item 2, WK1749, Standard.  The results of this study indicate that both stents exhibit high corrosion resistance in the PBS solution with stainless steel exhibiting slightly better corrosion resistance.  However, electropolished nitinol stents appeared to have better corrosion resistance as compared to the unpolished ones.

Summary: Cyclic Potentiodynamic polarization studies were conducted to compare the corrosion resistance of nitinol and stainless steel cardiovascular stents in Phosphate Buffered Saline (PBS) solution at 37 °C. Both stents exhibit high corrosion resistance in the PBS solution with stainless steel exhibiting slightly better corrosion resistance. However, electropolished nitinol stents appeared to have better corrosion resistance as compared to the unpolished ones.